AInstrument can be used to sense the conductivity of aqueous solutions continuously, and to add agent. Apparatus for remote measurments.
 2xS-F-K2A or …-K2AI 2xS-F-K2B or …-K2BI 2xS-F-K2C or …-K2CI 2xS-F-K4A or …-K4AI
|
 2xS-T-K2A or …-K2AI 2xS-T-K2B or …-K2BI 2xS-T-K2C or …-K2CI 2xS-T-K4A or …-K4AI
|
|
|
SPECIFICATION
|
|
Measure range:
…K2A
|
0...200μS/cm (1μS/cm...1mS/cm) k=1/cm 30Hz
0...20μS/cm (0,1μS/cm...100μS/cm) k=0,1/cm 30Hz
0...2μS/cm (0,01μS/cm...10μS/cm) k=0,01/cm 30Hz
|
|
…K2B
|
0...200μS/cm (1μS/cm...1mS/cm) k=1/cm 300Hz
0...20μS/cm (0,1μS/cm...100μS/cm) k=0,1/cm 300Hz
0...2μS/cm (0,01μS/cm...10μS/cm) k=0,01/cm 300Hz
|
|
…K2C
|
0...200mS/cm (200μS/cm...200mS/cm) k=1/cm 3kHz
0...20mS/cm (20μS/cm...20mS/cm) k=0,1/cm 3kHz
|
|
…K4A
|
0...200mS/cm (200-S/cm...200mS/cm) k=1/cm 3kHz
0...20mS/cm (20μS/cm...20mS/cm) k=0,1/cm 3kHz
0...2S/cm (2mS/cm...2S/cm) k=10/cm 3kHz
|
|
-Accuracy:
|
±0,2% ±1digit
|
|
Temperature compensation:
|
Tref=25°C-ra auto./manual
|
|
Measuring range -T:
|
-100-200 °C
|
|
Temperature sensor:
|
Pt100
|
|
- Accuracy:
|
±0,2 °C ±1digit
|
|
Current output:
|
0/4..20mA Rtmax=500 ohm
|
|
- 2. Current output :
|
2xS-..-KI (optional)
|
|
Adjustable output parameter:
|
K or T
|
|
The input is galvanically isolated from the output.
|
|
Limits outputs:
|
morse contact relay
|
|
- Minimum and maximum limits:
|
adjustable in the whole range
|
|
- Hysteresis:
|
adjustable
|
|
- Power supply:
|
230V AC 3A (Φ=1)
|
|
Digital out:
|
RS232 or RS485 (optional)
|
|
Power supply:
|
230V AC ±10% 50-60Hz 6VA, II. ÉVO
|
|
Ambient temperature:
|
-15 ..+45 °C
|
|
Protecting: -F- type
|
IP 65
|
|
..-T-.. type front-board
|
IP 44
|
|
..-T-.. type back-board
|
IP 20
|
|
- Size: ..-F-.. type
|
166x160x106mm
|
|
..-T-.. type
|
144x72x146mm
|
|
..-T-.. type cutout size:
|
138,3x67mm
|
|
- Weight:
|
approx.1Kg
|
|
- Electrodes: K
|
2 and 4 elektrodes sensor
|
|
T
|
Pt100
|
|

K2-34 Conductivity measuring sensor
|