Conductivity measuring

Conductivity measuring and controlling unit

 AInstrument can be used to sense the conductivity of aqueous solutions continuously, and to add agent. Apparatus for remote measurments.

 

 

 

SPECIFICATION
Measure range:

 

2xS.5-…-K2A

0…200μS/cm (1μS/cm…1mS/cm)   k=1/cm  30Hz
0…20μS/cm (0,1μS/cm…100μS/cm)  k=0,1/cm 30Hz
0…2μS/cm (0,01μS/cm…10μS/cm) k=0,01/cm 30Hz

 

2xS.5-…-K2B

0…200μS/cm (1μS/cm…1mS/cm)  k=1/cm 300Hz

0…20μS/cm (0,1μS/cm…100μS/cm)  k=0,1/cm 300Hz
0…2μS/cm (0,01μS/cm…10μS/cm) k=0,01/cm 300Hz

 

2xS.5-…-K2C

0…200mS/cm (200μS/cm…200mS/cm) k=1/cm 3kHz

0…20mS/cm (20μS/cm…20mS/cm) k=0,1/cm 3kHz
0…2S/cm (2mS/cm…2S/cm)    k=10/cm  3kHz

 

          …K4A

0…200mS/cm (200μS/cm…200mS/cm) k=1/cm 3kHz
0…200mS/cm (200μS/cm…200mS/cm) k=1/cm 3kHz
0…2S/cm (2mS/cm…2S/cm)    k=10/cm  3kHz
-Accuracy: ±0,2% ±1digit
Temperature compensation: Tref=25°C auto./manual
Measuring range -T: -100…+200 °C
Temperature sensor: Pt100
Accuracy: ±0,2 °C ±1digit
Current output: 0/4..20mA Rtmax=500 Ω
– 2. Current output: 2xS.5-..-KI (optional)
 Adjustable output parameter: K or T
The input is galvanically isolated from the output.
Limits outputs: morse contact relay
Minimum and maximum limits: adjustable in the whole range
Hysteresis: adjustable
–Power supply: 230V AC 3A (Φ=1)
Digital out: MODBUS RS232 or RS485 (optional)
Power supply: 230V AC ±10% 50-60Hz 6VA,II. ÉVO
Ambient temperature: -20..+50°C
Protecting:   -F-   type IP65
                     ..-T-.. type front-board IP 44
                     ..-T-.. type back-board IP 20
Size:       ..-F-.. type 166x160x106
               ..-T-.. type 144x72x146mm
               ..-T-.. type cutout size: 138,3x67mm
Weight: Approx.1Kg
Electrodes: K 2 and 4 elektrodes sensor
                       T Pt100 

 

TYING UP

 

K2-3/4 Conductivity measuring sensor

K2-3/4 Conductivity measuring sensor